Calibration range and CMC for flow velocity measurement instruments

Measurement rangeMeasurement uncertainty CMC*
(0,15 ÷ 1,5) m/s0,05 m/s
(1,5 ÷ 7) m/s–0,012 · v + 0,173 m/s
v m/s
(7 ÷ 40) m/s0,0083 · v + 0,0313 m/s
v m/s

* The Calibration and Measurement Capability (CMC) presented within the scope of accreditation, representing the expanded uncertainty at a confidence level of approximately 95%, is defined as the smallest measurement uncertainty that a laboratory can achieve during the routine calibration of objects. It is established based on uncertainty budget analyses conducted during calibrations using the most accurate measuring instruments available for calibration. CMC represents the smallest value of expanded uncertainty related to the calibration result (the obtained difference between the measured quantity value and the value indicated by the calibrated instrument) that can be provided on the calibration certificate. For different calibration conditions and different instruments, the expanded uncertainty stated on the certificate may be larger.

Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.