Calibrated instruments
The laboratory performs calibrations of electronic instruments, including:
- relative and differential pressure gauges and transducers,
- absolute pressure gauges and transducers.
Calibration method
Depending on the expected accuracy of the instrument, the adopted measurement procedures outline 3 calibration methods:
- The based method, which should be used for instruments with an expected expanded uncertainty (k = 2) of measurement U > 0.2% FS. In this method, calibration is conducted once at 6 pressure points for increasing and decreasing pressures. Zero error, repeatability, and transducer hysteresis are determined based on three repeated measurement series at 2 pressure points (typically 0 and 50% FS).
- The standard method, which should be used for instruments with an expected expanded uncertainty (k = 2) of measurement 0.05% FS < U < 0.2% FS. In this method, calibration is conducted once at 11 pressure points for increasing and decreasing pressures. Zero error, repeatability, and transducer hysteresis are determined based on three repeated measurement series at 4 pressure points (typically 0, 20, 50, and 80% FS).
- The full method, which should be used for instruments with an expected expanded uncertainty (k = 2) of measurement U < 0.05% FS. In this method, calibration is conducted at 11 pressure points for increasing and decreasing pressures in three measurement series. Zero error, repeatability, and transducer hysteresis are determined separately for each calibration point.
The choice of measurement method according to which the calibration will be conducted is up to the Customer. Upon request, the Laboratory provides assistance in selecting the appropriate calibration method based on the accuracy and application of the measurement instrument intended for calibration.
Calibration in the negative pressure range
We perform calibrations of differential pressure gauges as part of accreditation in the range of positive pressure (+) and negative pressure (-) from 0 to 200 kPa.
Calibration in the negative pressure range is performed by applying positive pressure (positive reference value) to the negative port of the calibrated measuring instrument (negative measured value).
The method of presenting calibration results in the vacuum pressure range is illustrated on a sample calibration certificate.